我院于 2002年引进 GE Hispeed DX/I单排螺旋 CT机器,设备运行较为正常,但是由于客观因素存在,-些软件或硬件方面的小故障也不可避免。以下是我们在使用过程中总结的-些故障处理经验与技巧,供同行参考与探讨。

1 故障-1.1 故障现象机器在使用过程中报错:Error:1 88448 Slice End Error/RawStore ErorEror:1 88448 Host:bhsct Process:RawStoreFile:RawStore.C Line:1893Captured Erors:Eror:188448Host:bhsct Process:RawStoreFile:RawStore.C Line:1 893 Slice End Error.B1.2 故障分析查看错误代码可以看到原始数据保存出错,原始数据盘无法正常存取数据,需要更换。

收稿日期:2013-04-161.3 故障处理更换元数据盘,重新 安装 OS ,application,Patch,option软件,机器恢复正常,故障解决。

2 故障二2.1 故障现象Jedi/OGP在扫描过程中出现通讯故障,故障描述为:Error:186436 Host:GIMCCT Process:tgp-outFile:tgp-erors.C Line:395 OGP Event Error。

Code :20-1030 -01 Eror detected by :OGP OGPdid not receive any reply from JEDI in response to Ro-tor ON/OFF requestEror:1 86436 Host:GIMCCT Process:tgp-outFile:tgp-erors.c Line:395 OGP Event Eror。

Code :20-1030-01 Eror detected by :OGPOGP did not receive any reply from JEDI in re-sponse to Rotor 0N/0FFrequest2.2 故障分析可能引起的原因有 JEDI和 OGP的连接电缆;AC/DC保险丝;AC/DC,LVPS,HEATER板;旋面盖时电机是否工作,灭菌器内有电机工作的声音,排除电机故障。打开后盖检查两个行程开关P,因按开盖键,电机都有工作,排除形成开关故障。再检查限位开关,发现锁扣未能勾住,为门钩底部有断裂。因此压下面盖时不能正常关准虑用电焊焊接门钩底部,门钩铁片有变形,并且角度不准确,而购买-个门钩较便宜,新安装上门钩,旋紧螺丝。重新测试 ,压下面盖,能够正常关盖。

医疗装备2013第7期总结:该款低温灭菌器对水质的要求较高,因此需定期对机器进行维护,及时的更换滤芯,定期做好维护。有些问题是操作不注意而出现的偶然,而有些如果能够很好的做好维护,能更好地减少机器故障,使灭菌器在工作中发挥更好的作用。并且多检查,多发现,其实许多问题都能够找到原因,得到及时有效的解决,并且更好的节省成本,做好维护工作77转阳极板。

2.3 故障处理首先打开扫描架,检查 JEDI和 OGP的连接电缆接头,用万用表测量电缆是否有中断的地方,经检测电缆及连接无问题。然后检查 AC/DC保险丝,经万用表测量后发现保险丝熔断,更换保险丝,重新开机测试后可以正常使用,观察-天后故障未再出现,彻底解决问题。

3 故障三3.1 故障现象机器无法正常启动,ORP和ccb通讯失败,扫描架复位失败。

Error:260101008 The ORP failed to connect withthe CCB sub-system in the allotted time. Check tosee if subsystem is alive. Timeout value:40000mSThis was followed by Error:244540 ImmediateRequest cannot be honored since the firmware is in theD0W N state。

3.3 故障处理首先清理 ccb板表面灰尘或者其它附属连接物。打开维修菜单,选择准直器测试,测试结果失败。更换-块全新的ccb板,做准直器测试,测试通过,ccb故障排除。

4 故障四4.1 故障现象scout,axial,helica均无法正常曝光,故障描述为 :Host:orpProcess:orp-subsysFile:gen- exee.c Line:2376Host:JediErmes#:2601 18185Exception Class:AbortSeverity :PrimaryFile :gen-exec.cLine# :2376Function :No System Function ReportedScan Type:Axial Scan:50103/8/6(50-0204)Filament permanent open circuit。

Issued by the heater board when an open has beendetected and 3 restartshave been tried without success within 1 00 ms。

784.2 故障分析可能引起的原因有球管灯丝开路;heater板和高压邮箱电缆;heater板;阴极高压电缆。

4.3 故障处理依次检查 heater和高压邮箱电缆;检查 heater板连接线;检查阴极高压电缆,均无异常,然后测量球管灯丝参数,发现灯丝无阻值,故应是球管损坏,更换球管后故障解决。

5 故障五5.1 故障现象病人检查无法曝光。故障描述为:Host:IRIS Process:tgp outFile:tgp ininput.C。

Line:334 XG Error phase is 3 code is 80.detailerror code :80 - 0902H Hardware eror :tube Fansupply error。

5.2 故障分析可能引起的原因有 115v球管风扇冷却电压 和旋转阳极板。

5.3 故障处理首先用万用表测量 115v球管风扇冷却电压,没有问题。然后检查旋转阳极板 DS6灯,红灯亮,更换旋转阳极板后,故障解决。

6 总结以上是我院GE单排 CT在使用过程中出现过的故障,从中可以发现,无论在软件还是在硬件出现故障时均会有不同类别的错误代码及含义提示。

查阅相关说明书,基本能够对故障代码含义及可能产生的原因有详细的说明,可以作为查找 CT故障的重要依据。只要平时善于积累经验,细致地分析检修,并与厂方工程师进行充分沟通交流,-些常见故障是自己完全可以解决的。