我院于 2002年引进 GE Hispeed DX/I单排螺旋 CT机器,设备运行较为正常,但是由于客观因素存在,-些软件或硬件方面的小故障也不可避免。以下是我们在使用过程中总结的-些故障处理经验与技巧,供同行参考与探讨。

Error:1 88448 Slice End Error/ RawStore ErorEror:1 88448 Host:bhsct Process:RawStoreFile:RawStore.c Line:1893Captured Erors:Eror:188448Host:bhsct Process:RawStoreFile:RawStore.C Line:1 893 Slice End Eror.B故障分析:查看错误代码可以看到原始数据保存出错,原始数据盘无法正常存取数据,需要更换。

收稿日期:2013-05-29故障处理:更换元数据盘,重新安装 OS,application,Patch,option软件,机器恢复正常,故障解决。

故障二:Jedi/OGP在扫描过程中出现通讯故障,故障描述为:Error:1 86436 Host:GIMCCT Process:tgp-outFile:tgp-errors.c Line:395 OGP Event Eror。

Code :20-1030-O1 Eror detected by :OGP OGPdid not receive any reply from JEDI in response to Ro-tor ON/OFF requestEror:1 86436 Host:GIMCCT Process:tgp-outFile:tgp-erors.c Line:395 OGP Event Error。

Code :20-1030-01 Error detected by :OGPOGP did not receive any reply from JEDI in re-sponse to Rotor 0N/0FFre0uest故障分析:可能引起的原因有 JEDI和 OGP的连接电缆;AC/DC保险丝;AC/DC,LVPS,HEATER板 ;旋带电、放电声音等立即关机、切断电源,向上级技师或科主任、器械师、维修人员报告,严禁带故障工作。防止故障进-步扩大造成严重后果。

5 定期维修安全检查与功能和工作可靠性密切相关,必须至少每两年执行-次安全检查,由专业工程师完成。

(1)目视检查设备的完整性以及有无明显损66坏或缺陷、污迹,卡住的部件和磨损情况 (所规定预防性保护)。

故障处理:首先打开扫描架,检查 JEDI和 OGP的连接电缆接头,用万用表测量电缆是否有中断的地方,经检测电缆及连接无问题。然后检查 AC/DC保险丝,经万用表测量后发现保险丝熔断,更换保险丝,重新开机测试后可以正常使用,观察-天后故障未再出现,彻底解决问题。

故障三:机器无法正常启动,ORP和 ccb通讯失败,扫描架复位失败。

Error:260101008 The ORP failed to connect withthe CCB sub-system in the allotted time. Check tosee if subsystem is alive.Timeout value:40000mSThis was followed by Error:244540 ImmediateRequest cannot be honored since the firmware is in theDOWN state。

故障处理:首先清理 ccb板表面灰尘或者其它附属连接物。打开维修菜单,选择准直器测试,测试结果失败。更换-块全新的 ccb板,做准直器测试 ,测试通过,ccb故障排除。

故障四:scout,axial,helica均无法正常曝光,故障描述为:Host:orp Process:orp-subsysFile:gen-exec.c Line:2376Host:JediErm es#:2601 18185Exception Class :Abort Severity :PrimaryFile:gen-exec.c Line# :2376Function :No System Function ReportedScan Type:Axial Scan:50103/8/6(50-O204)Filament perm anent open circuit。

Issued by the heater board when an open has beendetected and 3 restartshave been tried without success within 100 ms. Cz6,Z7 97h,Oh故障分析:医疗装备 2013第9期可能引起的原因有球管灯丝开路;heater板和高压邮箱电缆;heater板;阴极高压电缆。

故障处理:依次检查 heater和高压邮箱电缆;检查 heater板连接线;检查阴极高压电缆,均无异常,然后测量球管灯丝参数,发现灯丝无阻值,故应是球管损坏,更换球管后故障解决。

故障五:病人检查无法曝光。故障描述为:Host:IRIS Process:tgp out File:tgp in input.cLine:334 XG Error phase is 3 code is 80.detailelTor code :80 - 0902H Hardware eror :tube Fansupply eror故障分析:可能引起的原因有115V球管风扇冷却电压 和旋转阳极板。

故障处理:首先用万用表测量 115V球管风扇冷却电压,没有问题。然后检查旋转阳极板 DS6灯,红灯亮,更换旋转阳极板后,故障解决。

总结:以上是我院 GE单排 CT在使用过程中出现过的故障,从中可以发现,无论在软件还是在硬件出现故障时均会有不同类别的错误代码及含义提示。

查阅相关说明书,基本能够对故障代码含义及可能产生的原因有详细的说明,可以作为查找 CT故障的重要依据。只要平时善于积累经验,细致地分析检修,并与厂方工程师进行充分沟通交流,-些常见故障是自己完全可以解决的。